Spécifications

Aussi livrable comme "version complete" avec toutes les accessoires.

HDM97 BO - Testeur pour hémo dialyse

HDM97 BO Mesureur de conductivité, temperature et pression pour des appareils de dialyse - IBP Medical

Technical Data
     
Conductivity:  4 measuring ranges - Quadropole electrode
Temperature compensation is user-selectable depending on the brand of dialysis machine  
Range: 0 … 30,00 mS/cm
Accuracy: 0 … 200 uS/cm ± 0.6 uS/cm
          201 … 2000 uS/cm ± 6 uS/cm
          12 ... 16 mS/cm ± 0.03 mS/cm
          otherwise ± 0.06 mS/cm

With option C, the conductivity/temperature probe may be disconnected.
This option is available for all meters.
 
 
Temperature:   Sensor integrated in the conductivity electrode  
Range: 0...100 °C
Resolution: 0.01 °C
Accuracy: 25 ... 40°C ± 0.05 °C, otherwise ±  0.1 °C   
 
Pressure:    Reading selectable as mmHg, kPa, Bar or PSI
Version HDM97BQ, BP and BO
 
Range:    -700 to + 1900 mmHg, 2,5 Bar
Resolution:    0.1 mmHg 
Accuracy:    0 ... 300 mmHg ± 1 mmHg
otherwise ± 2 mmHg
    in temperature range 20 to 23 °C
Ceramic sensor fluid resistant

Option H with higher accuracy
Range:    -700 to + 1900 mmHg, 2,5 Bar
Resolution:    0.01 mmHg 
Accuracy:    0 ... 300 mmHg ± 0.5 mmHg,
301 ... 1000 mmHg ± 1 mmHg
otherwise ± 2 mmHg
    in temperature range 20 to 23 °C
Sensor not fluid resistent 
 
Power Supply:   Li-Ion-Polymer battery, rechargeable with 5 V DC, 500 mA
External charger:
Input Voltage 100 … 245 V / 50/60 Hz
Output Voltage 5 V DC, 500 mA
Or via USB-Interface from a computer
Operates up to 200 hours with fully charged battery
Dimensions and Weight  165 x 82 x 33 mm - 6.5" x 3.2" x 1.3”
0.4 Kg - 0.88 lbs including probe

  • CE logo