Spécifications
Aussi livrable comme "version complete" avec toutes les accessoires.
HDM97 BO - Testeur pour hémo dialyse
HDM97 BO Mesureur de conductivité, temperature et pression pour des appareils de dialyse - IBP Medical
Technical Data
Conductivity: 4 measuring ranges - Quadropole electrode
Temperature compensation is user-selectable depending on the brand of dialysis machine
Range: 0 … 30,00 mS/cm
Accuracy: 0 … 200 uS/cm ± 0.6 uS/cm
201 … 2000 uS/cm ± 6 uS/cm
12 ... 16 mS/cm ± 0.03 mS/cm
otherwise ± 0.06 mS/cm
With option C, the conductivity/temperature probe may be disconnected.
This option is available for all meters.
Temperature: Sensor integrated in the conductivity electrode
Range: 0...100 °C
Resolution: 0.01 °C
Accuracy: 25 ... 40°C ± 0.05 °C, otherwise ± 0.1 °C
Pressure: Reading selectable as mmHg, kPa, Bar or PSI
Version HDM97BQ, BP and BO
Range: -700 to + 1900 mmHg, 2,5 Bar
Resolution: 0.1 mmHg
Accuracy: 0 ... 300 mmHg ± 1 mmHg
otherwise ± 2 mmHg
in temperature range 20 to 23 °C
Ceramic sensor fluid resistant
Option H with higher accuracy
Range: -700 to + 1900 mmHg, 2,5 Bar
Resolution: 0.01 mmHg
Accuracy: 0 ... 300 mmHg ± 0.5 mmHg,
301 ... 1000 mmHg ± 1 mmHg
otherwise ± 2 mmHg
in temperature range 20 to 23 °C
Sensor not fluid resistent
Power Supply: Li-Ion-Polymer battery, rechargeable with 5 V DC, 500 mA
External charger:
Input Voltage 100 … 245 V / 50/60 Hz
Output Voltage 5 V DC, 500 mA
Or via USB-Interface from a computer
Operates up to 200 hours with fully charged battery
Dimensions and Weight 165 x 82 x 33 mm - 6.5" x 3.2" x 1.3”
0.4 Kg - 0.88 lbs including probe